User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Doric DS-100-T2:Strain Gage Transducer Indicator DATE: 15-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 211 STANDARD: Measurements Group 1550A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following Standard test equipment is required: 1.004 DISP 1. Measurements Group 1550A Strain Indicator Calibrator 1.004 DISP [32] DTB # 55-2 1.004 DISP 2. Strain Indicator Calibration Cable 1.005 STD Measurements Group 1550A 1.006 HEAD {* LINEARITY TEST *} 1.007 HEAD {Full scale = 12000} 1.008 JMP 2.001 1.009 EVAL 2.001 DISP Set the UUT front panel controls as follows: 2.001 DISP POWER .............................................. OFF 2.002 DISP With the UUT disconnected from AC power, pull out the 2.002 DISP plug-in board by pressing release catch located under 2.002 DISP the UUT front panel. 2.003 DISP Set the 1550A 350 Ohm bridge S+ shorting link to the 2.003 DISP closed position. 2.004 DISP Set the 1550A front panel controls as follows: 2.004 DISP MICROSTRAIN buttons ....................... 000 2.004 DISP CIRCUIT button ............................ 350 Ohm 2.004 DISP POLARITY button ........................... positive (+) 2.005 DISP Connect the Strain Indicator Calibration cable to the 2.005 DISP UUT rear panel connector. 2.006 DISP [32] Connect the 1550A and the UUT using the Strain 2.006 DISP [32] Indicator Calibration cable as follows: 2.006 DISP [32] [27][91]1m 1550A TO PLDT 2.006 DISP [32] P+ .........> P+ (orange & red) 2.006 DISP [32] P- .........> P- (black,blue & GND) 2.006 DISP [32] S+ .........> S+ (green & white) 2.006 DISP [32] S- .........> S- (white) 2.007 DISP Connect the UUT to AC power and turn the UUT on with 2.007 DISP the SAMPLE RATE control set to fully ccw (on position). 2.008 DISP Adjust the 1550A 350 Ohm ± ZERO control, while togling 2.008 DISP between the 1550A + and - POLARITY switch positions, 2.008 DISP for a UUT reading that is the same in magnitude for 2.008 DISP both positions. 2.009 DISP Set the 1550A front panel controls as follows: 2.009 DISP POLARITY button ........................... positive (+) 2.010 DISP Set the UUT mV/V switch to 4mV/V. 2.011 DISP Adjust the UUT ZERO control for a UUT reading of 0000. 2.012 DISP Set the 1550A front panel controls as follows: 2.012 DISP MICROSTRAIN PER STEP buttons ....................... 100 2.013 DISP Adjust the UUT SPAN (COARSE, FINE and FINE SPAN) 2.013 DISP controls for a UUT reading of 12000 counts. 2.014 DISP Set the 1550A front panel controls as follows: 2.014 DISP MICROSTRAIN PER STEP buttons ....................... 090 2.015 MEMI Use keyboard to enter actual UUT reading: 2.016 MATH MEM1 = 10800 2.017 ACC 10800µÆ 0.025% 1U 2.018 MEME 2.019 MEMC µÆ 3.86U #! WARNING: Test Tol 3.86, Sys Tol 3.7, TUR 1.043 (< 4.00). 3.001 DISP Set the 1550A front panel controls as follows: 3.001 DISP MICROSTRAIN PER STEP buttons ....................... 080 3.002 MEMI Use keyboard to enter actual UUT reading: 3.003 MATH MEM1 = 9600 3.004 ACC 9600µÆ 0.025% 1U 3.005 MEME 3.006 MEMC µÆ 3.62U #! WARNING: Test Tol 3.62, Sys Tol 3.4, TUR 1.065 (< 4.00). 4.001 DISP Set the 1550A front panel controls as follows: 4.001 DISP MICROSTRAIN PER STEP buttons ....................... 070 4.002 MEMI Use keyboard to enter actual UUT reading: 4.003 MATH MEM1 = 8400 4.004 ACC 8400µÆ 0.025% 1U 4.005 MEME 4.006 MEMC µÆ 3.38U #! WARNING: Test Tol 3.38, Sys Tol 3.1, TUR 1.090 (< 4.00). 5.001 DISP Set the 1550A front panel controls as follows: 5.001 DISP MICROSTRAIN PER STEP buttons ....................... 060 5.002 MEMI Use keyboard to enter actual UUT reading: 5.003 MATH MEM1 = 7200 5.004 ACC 7200µÆ 0.025% 1U 5.005 MEME 5.006 MEMC µÆ 3.14U #! WARNING: Test Tol 3.14, Sys Tol 2.8, TUR 1.121 (< 4.00). 6.001 DISP Set the 1550A front panel controls as follows: 6.001 DISP MICROSTRAIN PER STEP buttons ....................... 050 6.002 MEMI Use keyboard to enter actual UUT reading: 6.003 MATH MEM1 = 6000 6.004 ACC 6000µÆ 0.025% 1U 6.005 MEME 6.006 MEMC µÆ 2.90U #! WARNING: Test Tol 2.9, Sys Tol 2.5, TUR 1.160 (< 4.00). 7.001 DISP Set the 1550A front panel controls as follows: 7.001 DISP MICROSTRAIN PER STEP buttons ....................... 040 7.002 MEMI Use keyboard to enter actual UUT reading: 7.003 MATH MEM1 = 4800 7.004 ACC 4800µÆ 0.025% 1U 7.005 MEME 7.006 MEMC µÆ 2.66U #! WARNING: Test Tol 2.66, Sys Tol 2.2, TUR 1.209 (< 4.00). 8.001 DISP Set the 1550A front panel controls as follows: 8.001 DISP MICROSTRAIN PER STEP buttons ....................... 030 8.002 MEMI Use keyboard to enter actual UUT reading: 8.003 MATH MEM1 = 3600 8.004 ACC 3600µÆ 0.025% 1U 8.005 MEME 8.006 MEMC µÆ 2.42U #! WARNING: Test Tol 2.42, Sys Tol 1.9, TUR 1.274 (< 4.00). 9.001 DISP Set the 1550A front panel controls as follows: 9.001 DISP MICROSTRAIN PER STEP buttons ....................... 020 9.002 MEMI Use keyboard to enter actual UUT reading: 9.003 MATH MEM1 = 2400 9.004 ACC 2400µÆ 0.025% 1U 9.005 MEME 9.006 MEMC µÆ 2.18U #! WARNING: Test Tol 2.18, Sys Tol 1.6, TUR 1.362 (< 4.00). 10.001 DISP Set the 1550A front panel controls as follows: 10.001 DISP MICROSTRAIN PER STEP buttons ....................... 010 10.002 MEMI Use keyboard to enter actual UUT reading: 10.003 MATH MEM1 = 1200 10.004 ACC 1200µÆ 0.025% 1U 10.005 MEME 10.006 MEMC µÆ 1.94U #! WARNING: Test Tol 1.94, Sys Tol 1.3, TUR 1.492 (< 4.00). 11.001 DISP Set the 1550A front panel controls as follows: 11.001 DISP MICROSTRAIN PER STEP buttons ....................... 009 11.002 MEMI Use keyboard to enter actual UUT reading: 11.003 MATH MEM1 = 1080 11.004 ACC 1080µÆ 0.025% 1U 11.005 MEME 11.006 MEMC µÆ 1.92U #! WARNING: Test Tol 1.92, Sys Tol 1.27, TUR 1.512 (< 4.00). 12.001 DISP Set the 1550A front panel controls as follows: 12.001 DISP MICROSTRAIN PER STEP buttons ....................... 008 12.002 MEMI Use keyboard to enter actual UUT reading: 12.003 MATH MEM1 = 960 12.004 ACC 960µÆ 0.025% 1U 12.005 MEME 12.006 MEMC µÆ 1.89U #! WARNING: Test Tol 1.89, Sys Tol 1.24, TUR 1.524 (< 4.00). 13.001 DISP Set the 1550A front panel controls as follows: 13.001 DISP MICROSTRAIN PER STEP buttons ....................... 007 13.002 MEMI Use keyboard to enter actual UUT reading: 13.003 MATH MEM1 = 840 13.004 ACC 840µÆ 0.025% 1U 13.005 MEME 13.006 MEMC µÆ 1.878U #! WARNING: Test Tol 1.878, Sys Tol 1.21, TUR 1.552 (< 4.00). 14.001 DISP Set the 1550A front panel controls as follows: 14.001 DISP MICROSTRAIN PER STEP buttons ....................... 006 14.002 MEMI Use keyboard to enter actual UUT reading: 14.003 MATH MEM1 = 720 14.004 ACC 720µÆ 0.025% 1U 14.005 MEME 14.006 MEMC µÆ 1.84U #! WARNING: Test Tol 1.84, Sys Tol 1.18, TUR 1.559 (< 4.00). 15.001 DISP Set the 1550A front panel controls as follows: 15.001 DISP MICROSTRAIN PER STEP buttons ....................... 005 15.002 MEMI Use keyboard to enter actual UUT reading: 15.003 MATH MEM1 = 600 15.004 ACC 600µÆ 0.025% 1U 15.005 MEME 15.006 MEMC µÆ 1.82U #! WARNING: Test Tol 1.82, Sys Tol 1.15, TUR 1.583 (< 4.00). 16.001 DISP Set the 1550A front panel controls as follows: 16.001 DISP MICROSTRAIN PER STEP buttons ....................... 004 16.002 MEMI Use keyboard to enter actual UUT reading: 16.003 MATH MEM1 = 480 16.004 ACC 480µÆ 0.025% 1U 16.005 MEME 16.006 MEMC µÆ 1.80U #! WARNING: Test Tol 1.8, Sys Tol 1.12, TUR 1.607 (< 4.00). 17.001 DISP Set the 1550A front panel controls as follows: 17.001 DISP MICROSTRAIN PER STEP buttons ....................... 003 17.002 MEMI Use keyboard to enter actual UUT reading: 17.003 MATH MEM1 = 360 17.004 ACC 360µÆ 0.025% 1U 17.005 MEME 17.006 MEMC µÆ 1.77U #! WARNING: Test Tol 1.77, Sys Tol 1.09, TUR 1.624 (< 4.00). 18.001 DISP Set the 1550A front panel controls as follows: 18.001 DISP MICROSTRAIN PER STEP buttons ....................... 002 18.002 MEMI Use keyboard to enter actual UUT reading: 18.003 MATH MEM1 = 240 18.004 ACC 240µÆ 0.025% 1U 18.005 MEME 18.006 MEMC µÆ 1.75U #! WARNING: Test Tol 1.75, Sys Tol 1.06, TUR 1.651 (< 4.00). 19.001 DISP Set the 1550A front panel controls as follows: 19.001 DISP MICROSTRAIN PER STEP buttons ....................... 001 19.002 MEMI Use keyboard to enter actual UUT reading: 19.003 MATH MEM1 = 120 19.004 ACC 120µÆ 0.025% 1U 19.005 MEME 19.006 MEMC µÆ 1.72U #! WARNING: Test Tol 1.72, Sys Tol 1.03, TUR 1.670 (< 4.00). 20.001 DISP For DTB# 55-4 set the mV/V switch to 2mV/V. 20.001 DISP 20.001 DISP For DTB# 55-5 set the mV/V switch to 3.8mV/V. 20.002 END #! T.U.R.s less than 4.00: 18 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.